Colorlite Diagnostic and Correction Process Using the Mosaic Test
Mosaic Test for Optics
Rx form (for recording measurement and lens data)
The Mosaic Test helps diagnose red-green color vision deficiency and identify the most effective Colorlite corrective lens. We recommend also using the Ishihara Test, as it is the official “gold standard” and patients often want to know whether they would pass it. The Mosaic test works as well as the Colorlite color blindness diagnostic book on a properly set-up monitor.
Mosaic Test Overview
A smaller square (mosaic) must be identified within a larger square. Each tiny mosaic pulses and changes brightness. The test has 20 difficulty levels that diagnose red-green, Deutan (purple-green) and Protan (purple-red) color vision deficiency.
Monitor Setup
- Do not use mobile phones. Use a monitor or tablet.
- Remove all colored lenses, unless re-evaluating with Colorlite glasses.
- Set screen brightness to maximum (minimum 75%).
- Disable all color filters, such as night mode, blue light filters, or colorblind modes.
- Avoid direct light on the screen; perform the test in a dark environment (except during adaptation of the Colorlite lens).
- Ensure monitor contrast is at least 75% (and set the color temperature to ‘warm’).
Testing Process
- If the patient wears glasses, put the necessary correction lenses into a universal trial frame.
- Perform the Ishihara Test and record the result (use the Rx form).
- Start the Mosaic Test by clicking the “Diagnostic” Continue until the diagnosis and recommended Colorlite lenses are shown.
- Put the first recommended pair of Colorlite lenses into the trial frame and allow the patient at least 2 minutes to adaptation. Tip: Adaptation is faster in a well-lit white enviroment.
- Click “Reassessment” to evaluate the Colorlite lenses and record the result (use the Rx form).
- Continue until you get a normal result (or close to normal) with one of the lenses.
- Also perform the Ishihara test with the best lens.
Guidelines
- You do not need to test all lenses once the best one gives a normal result (both on Mosaic and Ishihara tests), but at least 3 different types should be tried.
- If two lenses give similar results, choose the less tinted one (lower id number).
Example: D10 is preferred over D15. - For each lens, allow a minimum of 2 minutes for chromatic adaptation. Proper adaptation is when white paper looks white, not colored. Adaptation is an essential part of the correction!